R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.13 Calibration
2-63
2.2.13.2 Total Gain
Prior to performing this calibration, be sure to set the RBW, dB/div and reference level as the measure-
ment conditions.
Setup
Connect the calibration signal.
1.
Connect the N-BNC adapter to the INPUT connector on the front panel.
2.
Connect the Input cable between the CAL OUT and INPUT connectors on the
front panel.
Setting measurement conditions
3.
Set the RBW, dB/div and the reference level to the values which are actually used
in measurements.
Performing the calibration
4.
Press SHIFT and 7(CAL).
The menu used for calibration appears (See Figure 2-72).
5.
Press Total Gain.
Calibration is performed using the current measurement conditions.
CAUTION:
Perform the Total Gain calibration again if you have changed the RBW,
dB/div and reference level after completing the Total Gain calibration.