
2. Spray with water and/or detergent, using a soft brush to dislodge any particulate
matter.
3. Visually inspect the electrode for signs of damage.
4. Calibrate the electrode.
5. Replace the PTFE tape and re-mount into the system, taking care to avoid
torsion on the cable.
Slow response or non-reproducible measurements are signs that the electrode has
become coated or clogged.
The pH glass is susceptible to mounting by many substances. The speed of response,
normally 95% of the reading in less than 10 seconds, is dramatically degraded when the
pH glass is coated.
For pH electrode, to restore the speed of response, clean the bulb with a high quality
detergent, methyl alcohol or other suitable solvent using a "Q-tip". Rinse well with
distilled water and retest. If the electrode now responds, but erratically, soak the sensor
in 0.1 Molar HCl for 5 minutes. Remove and rinse with water and place in 0.1 Molar
NaOH for 5 minutes. Remove, rinse again and then place the sensor in pH 4. buffer for
10 minutes before use.
ORP/REDOX electrode may be gently abraded by use of 600 grade wet silicone carbide
paper, jewelers rouge or very fine steel wool. Try one of the previous cleaning methods
first before using this method.
21