2.4 Peak
handling
The unit continuously check the actual GROSS load value against the acquired peak load value and
if the actual value is higher the peak load value will be updated.
Peak load value is showed on auxiliary display.
2.4.1 Peak load value indication
Pressing the key
9
the unit shows the value of the peak load acquired.
The value is updated continuously in case of load value increase.
2.4.2 Peak value printout
The A200 can be programmed to print the peak load value on a 24 column printer.
Printing is sent on the primary serial line if it is programmed as “DE” and can be initiated pressing
the “
E
” on keyboard or by closure of an external input programmed as “PRINT”.
Printout format is as follows:
18-11-99 11:09
Date and time
Peak: 5678 Kg
Peak load value
----------------------
Separator
2.4.3 Peak
reset
Peak value can be reset by:
•
by closure of an external input programmed as “RESET”.
•
by serial line with ADOS protocol (“m” command)
•
by Modbus protocol (E Coil)
•
Using the keyboard
To reset the peak value from keyboard the operator must press the
>0<
key.
The unit will prompt for confirmation:
O K ? ?
blinking
And the operator can confirm pressing the “
E
” key.