50
•
Operation Theory
Trigger Level digital setting
Trigger voltage
0xFF
9.92V
0xFE
9.84V
---
---
0x81
0.08V
0x80
0
0x7F
-0.08V
---
---
0x01
-9.92V
Table 11: Analog trigger SRC1 (EXTATRIG) ideal transfer
characteristic
The trigger signal is generated when the analog trigger condition is satis-
fied. There are five analog trigger conditions in the DAQ/PXI-20XX. The
DAQ/PXI-20XX uses 2 threshold voltages, Low_Threshold and
High_Threshold to build the 5 different trigger conditions. Users could
configure the trigger conditions easily by software.
4.5.2.1 Below-Low analog trigger condition
Figure 38 shows the below -low analog trigger condition, the trigger signal
is generated when the input analog signal is less than the Low_Threshold
voltage, and the High_Threshold setting is not used in this trigger condi-
tion.
Figure 38: Below-Low analog trigger condition