General safety rules
The EMC performance of this product is obtained only if the installation is
carried out in compliance with the applicable EMC rules. In particular, in
environments liable to emit strong EM disturbances, you must
- use shielded cables and connections for the interfaces
- protect the power line by a shielding from the power source to within
3 metres of the power input to the product.
The enclosure is designed to ensure normal safety conditions in its normal
operating environment (in a rack). When used on a table, or when
handling the enclosure make sure no objects are introduced into the
ventilation holes and that these holes are not blocked.
When switching off equipment containing condensers charged at more
than 60 VDC or 25 VAC, take all necessary precautions to protect against
contact with the pins on the connectors (single phase motors, equipment
with mains filter, frequency inverter, monitor, etc.)
Before switching on the power to this controller, the user must be familiar
with this manual and must follow the safety instructions herein.
• The Controllers must be connected to an electrical installation fitted
with a type TT earthed socket in compliance with ruling 88-1056 of 14th
November, 1988 (for French regulation).
• Our products are designed to meet all CEE regulations in effect. Any
modifications by the user are liable to break compliance with regulations
and jeopardise the EMC (Electromagnetic Compatibility) performance
and product safety. The manufacturer renounces all responsibility for the
consequences of any such intervention.
Safety instructions for the
installation