Detailed description
of RS232 commands
Command
Quer
y
Set
Syntax
Reply
Description
TEE
$TEE
$a
Test EEPROM
a= result of EEPROM test
PASS -> OK
ERR_ -> error
TID
$TID
$a
Query type of gauge connected
a = gauge connected
AP, ACC, AHC, AD, or NoGau
TIO
$TIO,a
$OK
Test set point relays
a = state of set point relays
0 -> initial state
1 -> set points and 2 OFF
2 -> set points and 2 ON
$TIO,a,b
$OK
TKY
$TKY
$a
Test panel buttons
a= status of command buttons
x*** -> SET button
*x** -> ENTER button
**x* -> ^ button
***x -> v button
0 for OFF, 1 for ON
TRS
$TRS
$OK
Test RS232 link
End of test when ‘ETX’ ( ‘Ctrl C’ ) is transmitted..
UNI
$UNI,?
$a
Set pressure measurement units
a = pressure measurement units
0 -> Pa
1 -> Torr
2 -> mbar
$UNI,a
$OK
VER
$VER
$a
Query controller software version
a= software version