GB 03435 - Edition Edition 03 - October 09
13/14
A Commissioning
B Product
description
C Settings
D Test
mode
E RS232
mode
F Appendices
General settings for the
ACS 2000 controller
Restoring factory
settings
Description
This setting causes the ACS 2000 controller to reset all settings their
default values (factory default setting).
Accessing this setting:
Access:
SET/PROG2/6. FACT
- Select «ON» and press
ENTER
.
Factory default setting
See appendix 5.
Available Remote
Inputs/Output
Analog recording output
A 0-10 V non configurable analog output is available at the Remote
connector (pins 2 & 3). Its value is independent of the type of gauge
connected to the channel.
The signal can be used by applying the following formula:
output (V) = 1/2 (Log P
disp
+ 12)
or
P
disp
= 10
(2 x output voltage(V) – 12)
(P
disp
: value of displayed pressure)
See Appendix 4 for wiring of the Remote connector.
A voltage of 10V, 9.5V, or 0.5V is considered an error.
See list of error messages in Appendix 6.