GB 03435 - Edition Edition 03 - October 09
11/14
A Commissioning
B Product
description
C Settings
D Test
mode
E RS232
mode
F Appendices
General settings for the
ACS 2000 controller
Settings for
measurements
Digit
Units
Description
This setting lets you chose between 2 or 3 digit display for values.
Accessing this setting:
Access:
SET/PROG2/1. DIGIT
Changing this setting does not require the ACS2000 controller to be
restarted.
Factory default setting
The pressure display is factory set to
2
digits.
This setting concerns all gauges except the capacitive gauge.
Description
This setting lets you choose between the following units for displaying
pressure:
- mbar
- Torr
- Pascal
The pressure units are shown in the bottom right of the screen.
Accessing this setting:
Access:
SET/PROG2/3. UNIT
Changing this setting does not require the ACS2000 controller to be
restarted.
Factory default setting
The default pressure unit is factory set to
mbar
.