110
If all the tests under the current fixed point are completed, it will return to the previous interface to select
another fixed point or DUT to continue testing.
8. When all test points of all DUT are finished, Click
to enter the save interface. Click
to modify the
set point used to calculate the coefficient. Click
to save the data to the USB flash drive, click
to save the
data locally.
Comparison Method
Refer to the platinum RTD calibration process, but note that the SPRT comparison method can select to enable
the self-heating effect test and the zero power tests.
3) Thermocouple Calibration
The thermocouple calibration process can refer to the platinum RTD calibration process, but the
thermocouple calibration requires the thermocouple type to be selected first and the reference sensor is
selected next. Pay attention to the cold junction type setting when configuring the thermocouple. There is no
coefficient calculation and standard resistance function for thermocouple calibration.