64
channel
Wiring system
2-, 3-, or 4-wire
2-, 3-, or 4-wire configurations for each channel
Sensor type
Humidity sensor, pressure sensor
When in channel of 02~10, the sensor types can be selected.
CJC Type
Internal/External
Internal will use the interal CJC temperature measurement. External
will use the temperature of the fixed value.
Fixed value
Manually input(-20-100)
℃
Choosing “thermocouple “to display Cold junction fixed value, the
default cold junction value is 0al will use the temperaturen
(-20-100)
℃
Open Circuit Detect
ON/OFF
Only for 01 channel , choosing relative measuring items of
"thermocouple" it is displayed that the connection of sensor is normal
or not
Wiring length
Manually input(0-50)m
Only for 01 channel, Set the wiring length of sensor
Wiring Settings
Front Panel
Click
to enter front panel settings, the external connecting cold junction channel of thermocouple can be
set.
Temperature Scanner