38
* Automatic current reversal.
PRT Accuracy Rating of Temperature
ADT286
Scanning
Speed
Temperature
24 Hour / °C
(23 ±1) °°C
90 Days / °C
(23 ± 5) °°C
1 year / °C
(23 ± 5) °°C
Temperature
Coefficient
°C/°C
Slow Speed
-200 °C
0.0005
0.0008
0.0008
0.0002
0 °C
0.0008
0.0031
0.0038
0.0008
300 °C
0.0018
0.0089
0.0089
0.0018
600 °C
0.0029
0.0146
0.0146
0.0030
Medium
Speed
-200 °C
0.0013
0.0016
0.0016
0.0002
0 °C
0.0014
0.0031
0.0038
0.0008
300 °C
0.0020
0.0089
0.0089
0.0018
600 °C
0.0029
0.0146
0.0146
0.0030
Fast
-200 °C
0.0039
0.0043
0.0043
0.0006
0 °C
0.0044
0.0047
0.0047
0.0013
300 °C
0.0053
0.0093
0.0093
0.0024
600 °C
0.0059
0.0152
0.0152
0.0036
*
The indicator is based on the electrical accuracy of the 4-wire PT100 PRT and does not include the accuracy of the PRT itself.
*Temperature maximum Resolution is 0.0001 °C.