62
Sheet 9 channel settings of temperature scanner
Settings
Operation
Explanation
Enabled
ON/OFF
Whether to display the measurement data
Label
Label
Customizable naming for the measurement channel
Function
Function setting
Including voltage, resistance, TC, standard TC, RTD, SPRT, Custom
RTD, Thermistor
Fixed delay/Extra delay
Manually input (0~60000)ms
The time it takes to switch from the other channel to the current
channel
High Resistance
ON/OFF
Improving
measurement
accuracy for
small-range
voltage
measurements
Wire
2-,3-, or 4-wire
For 2-, 3-, or 4-wire configuarations
Open Circuit Detect
ON/OFF
Choosing the relative measuring items of "thermocouple" to
indicate that connection is normal or not.
CJC Type
Internal/External/Custom
Internal will use the interal CJC temperature measurement. External
will use the temperature of the fixed value. Custom CJC will use the
reading of the other channel to be the CJC value for the
thermocouple being measured. The CJC channel is designated by
selectiong the scanner settings
and designating a CJC
channel. Example: REF 1 has a Type S TC with a cold junction in an
ice bath. CJC type is set to custom and REF 2 is the designated
channel. REF 2 has a SPRT selected and the SPRT is in the ice
bath. When measurements are made for the Type S TC, it will use
Содержание ADT286
Страница 1: ......
Страница 20: ...19 1 3 Base Structure Picture 1 ADT286 Mainframe...
Страница 22: ...21 Picture 2 Connection of inserting and unplugging of embedded scanners...
Страница 23: ...22 1 3 1 ADT286 TS Multi channel temperature signal scanner Picture 3 Temperature Device...
Страница 25: ...24 1 3 2 ADT286 PS Multi channel process signal scanner Picture 4 Process Device...
Страница 27: ...26 1 3 3 ADT286 Dock Remote intelligent wiring device Picture 5 Remote intelligent wiring device...
Страница 129: ......