start reading, or manually click
to skip the stabilization phase for direct reading.
9. After completing all the set point select
to enter the save interface to the internal memory of the
ADT286. Click the list of DUT sensors on the left to view the corresponding test results. Click
to select
the set point used for SPRT or CVD coefficient calculation and edit coefficient calculation for thermal
resistance. Click
to save the data to the USB flash drive, click
to save the data locally, and fill in the
file name, operator and notes when saving.
2) Standard Platinum Resistance Calibration:
Fixed points method and comparison method
Fixed points method
1. Click
to enter DUT type interface, click the standard platinum resistance item, and select fixed point
as test method.
2. Select the DUT sensor interface and select the positive and negative temperature calculation
coefficients to increase or decrease the DUT sensor (when the multiple sensors are inspected, the required