107
needs to be before measurements can be made.
Stability time: time fram over which the stability tolerance must be meet.
Additonal waiting time: additional time for stabilization
Channel scan cycles: number of times all channels are scanned
Sample number: number of readings at each channel
Channel scan round trip enable: When enabled the scan will be measuring increasing and then
decreasing (e.g. 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 10, 9, 8, 7, 6, 5,4,3,2,1) When round trip is enabled each
duplicate channel measurement will be averaged. This is useful when using a heatsource that is not
as stable as desired. Disabled round trip (1, 2, 3, 4, 5, 6, 7, 8, 9, 10)
6. Click
to enter the test configuration information settings
7. Click
to save the previous configuration and enter the pre-run test interface.
8. Click
to enter the probe calibration interface. In this interface, you can add the temperature source,
click
to start reading, the device will automatically control the temperature source, judge the stability and
Содержание ADT286