3-8
12701A Test Fixture Operation Manual
3.2
Resistance Measurement
Connecting the relay terminals with the test box
8.
Prepare the 2x2 dia.-1 dia. coaxial cable connection lead
(A01228-15).
Connect the tip of the connection lead to the relay terminal
FORCE/SENSE DG 1. Connect the relay terminal FORCE/SENSE
HI 1 and the test box 1 using the connection lead.
9.
Prepare the 2x2 dia.-1 dia. coaxial cable connection lead
(A01228-15).
Connect the tip of the connection lead to the relay terminal
FORCE/SENSE DG 2. Connect the relay terminal FORCE/SENSE
HI 2 and the test box 2 using the connection lead.
10.
Prepare the 2 dia.-2 dia. connection lead (A01215-15).
Connect the relay terminal FORCE/SENSE DG 2 and the TEST
BOX GUARD.
11.
Prepare the 2 dia.-1 dia. connection lead (A01214-20).
Connect the test box 3 onto the plug where the connection lead
(A01215-15) is connected to the relay terminal FORCE/SENSE DG
1, using the connection lead (A01214-20).
12.
Prepare the 2 dia.-1 dia. connection lead (A01214-20).
Connect the relay terminal FORCE/SENSE HI 3 and the test box 4.
The cables have been connected as shown in Figure 3-7.