SETTING TREATMENT OUTPUTS:
Once the protocol has been selected, the parameters adjusted per treatment requirements, and the electrodes placed
on the patient, the treatment outputs can be adjusted per the directions below
Once the parameters are adjusted, press the START/STOP button once. The display, depending on the length of
time lapsed from parameter adjustment to pressing the START/STOP button, will show one of the two following
screens:
1. If the START/STOP button is pressed immediately after parameter adjustment, the screen will show the protocol
selected (in this example, ANALGESIA NERVE BLOCK). If this screen is displayed, press START/STOP again to
get to the second screen (#2 below).
2. If the START/STOP button is pressed after a few seconds delay following modifying of the protocols (5 or more
seconds), or after pressing START/STOP again after the screen in #1 above, the following screen will be displayed.
NOTE:
At this point, the protocol parameters can be adjusted again, if desired. To adjust the parameters, press the
SELECT PARAMETERS button until the desired parameter to be modified is displayed. Press the TIME
PARAMETER UP or DOWN arrows to modify that parameter.
ANALGESIA NERVE BLOCK
LOCAL ACUTE
SET OUTPUTS A – B:
A: 0mA B: 0mA A