28 OMNISTIM
®
500A USER MANUAL
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
ANALGESIA – NERVE BLOCK:
Blocks pain by causing a temporary nerve block through reactive depolarization (Conduction block) of the pain
signal on its way to the spinal input. Also known as Wedensky inhibition. Duration of relief is typically from 1 to 2
hours. Faster onset of relief usually within 5 to 10 minutes.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
IFC Mode:
Full Field
Carrier frequency:
5.0 KHz
5.0 KHz or 10.0 KHz
Burst frequency:
Burst rate scan:
Burst scan time:
Beat:
Sweep:
Rate Scan Time:
Rate scan:
Vector:
OFF
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
10%
0 – 20%
Treatment time:
10 minutes
0 – 99 Minutes
Waveform:
Alternating Current
Output:
0 mA
0 – 70 mA
IFC CUSTOM (FULL FIELD):
Protocol fully adjustable by therapist, which can be customized for surface to deep tissue treatment.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
IFC Mode:
Full Field
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
80 Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
10 seconds
0 – 20 Seconds
Beat:
Sweep:
Rate Scan Time:
Rate scan:
ON On
or
OFF
Vector:
OFF
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
20%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA