OMNISTIM
®
500A USER MANUAL
33
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
HIGH PERFORMANCE MFAC RE-ED POWER:
Activation of muscle with electrical stimulation at a high intensity for a short time with short ON/OFF ramps
reduces atrophy and improves power. The patient should participate to the extent possible by contracting during the
stimulation ON time.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
2.5 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
70 Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
4 seconds
0 – 20 Seconds
Rate scan:
ON
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
10 minutes
0 – 99 Minutes
Channel B delay:
.0 seconds
.0 – 9.9 seconds
Ramp up:
0.2 seconds
.0 – 9.9 seconds
Ramp down:
1 seconds
.0 – 9.9 seconds
Time on:
10 seconds
1 – 30 seconds
Time off:
50 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
B DELAYED, ALTERNATE
or SIMULTANEOUS
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA
NMES SEQUENTIAL RE-ED:
Activation of muscle causing contraction of the muscles in a sequential manner. May be used to retrain functional
activity and reduce disuse atrophy in activities such as gait training for activation of swing phase to heel strike.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
2.5 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
70 Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
4 seconds
0 – 20 Seconds
Rate scan:
ON
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Channel B delay:
0.2 seconds
.0 – 9.9 seconds
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
10 seconds
1 – 30 seconds
Time off:
50 seconds
0 – 199 seconds
Channels A/B:
B DELAYED
B DELAYED, ALTERNATE
or SIMULTANEOUS
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA