42 OMNISTIM
®
500A USER MANUAL
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
INSPECT BATTERY CONTACTS / VERIFY BATTERY SWITCH SET TO RECHARGEABLE
1.
Remove batteries if they are installed.
2.
Inspect the battery contacts for any corrosion or discoloration. If corrosion or discoloration is found, perform the
following steps to clean the metal contacts:
•
Remove the power cord from the stimulator unit.
•
Hold the metal contact plate in place with one finger.
•
Use a small piece of sand paper to clean the surface of the metal contact thoroughly as needed. (Figure 3).
Figure 3
3.
Verify the switch inside the battery compartment is set to the RECHARGEABLE BAT. Position (assuming
rechargeable batteries are being used, which is recommended).
4.
Install the rechargeable batteries into the unit.
5.
Verify the battery door opens and closes correctly. The door should lock into place when closed.
INSPECT POWER SUPPLY OPERATION
1.
Connect the power cord to the power supply and then to a power outlet.
2.
Connect the power supply cable to the stimulator unit (Figure 4).
Figure 4
Cleaning the battery
contact (piece of sand
paper shown)
Connect
power supply
cable to
stimulator