52 OMNISTIM
®
500A USER MANUAL
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
Once the desired protocol is selected (in the example below, the selection is ANALGESIA – SENSORY), press the
START/STOP button. The following screen is displayed:
Adjust the output intensity by using the UP or DOWN buttons for OUTPUT A and OUTPUT B, as shown below:
When the desired output intensity is selected based on patient sensation, press the START/STOP button to initiate
the treatment. The following screen will be displayed. The treatment timer has started to count down at this point
(top right part of the screen). The protocols are displayed on the top line. Second line displays the treatment output
levels for both channels as well as the ON/OFF time operation (bars on the right side of display).
The timer in the upper right portion of the screen will continue to count down the treatment duration. Any
parameters, including output intensity and treatment time, can be adjusted during the treatment.
NOTE:
Special care should be taken if parameters are adjusted during treatment. Output intensity on both
channels should be decreased first prior to adjusting other parameters in order to prevent sudden changes that may
result in sudden burst of output to the patient.
Treatment will stop automatically when the treatment time counts down to zero. Treatment can also be stopped at
any time by pressing the START/STOP button, or thru the use of the Patient Safety Switch. Once the treatment is
completed, remove the electrodes from the patient and disconnect the lead wires from the electrodes. At this point,
the unit can be turned off by pressing the power button.
SET OUTPUTS A – B:
A: 0mA B: 0mA A
SET OUTPUTS A – B:
A: 7mA B: 7mA A
IFCff CONT 100Hz T 15
A: 7mA B: 7mA
A