ANALGESIA – SENSORY (FULL FIELD):
Target tissue is superficial and deep. Sensory stimulation activates A-beta fibers causing the release of spinal
Enkephalin and Dynorphin to block pain at the segmental level (Gate Control). Duration of relief is typically from
30 minutes to 2 hours. Fast onset of relief usually within 20 minutes.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
IFC Mode:
Full Field
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
100Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
10 seconds
0 – 20 Seconds
Beat:
Sweep:
Rate Scan Time:
Rate scan:
ON
ON or OFF
Vector:
Fast 90º
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
10%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA
ANALGESIA – SENSORY (FREQUENCY DIFFERENCE):
Target is deep tissue. Sensory stimulation activates A-beta fibers causing the release of spinal Enkephalin and
Dynorphin to block pain at the segmental level (Gate Control). Duration of relief is typically from 30 minutes to 2
hours. Fast onset of relief usually within 20 minutes.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
IFC Mode:
Frequency Difference
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
Burst rate scan:
Burst scan time:
Beat:
80 Hz
.0 – 250 Hz
Sweep:
120 Hz
.0 – 250 Hz
Rate Scan Time:
15 Seconds
0 – 20 Seconds
Rate scan:
Vector:
Fast 90º
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
0%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA