NMES – R.O.M.:
Activation of muscle causing contraction of the agonist and antagonist to move the joint and improve the R.O.M.
The patient should participate to the extent possible by contracting during the stimulation time.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Phase duration:
200 µS
30 µS – 500 µS
Pulse rate::
35 Hz
.0 Hz to 999 Hz
Phase duration mod:
Modulation time:
Modulation:
OFF
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
30 minutes
0 – 99 minutes
Channel B delay:
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
4 seconds
1 – 30 seconds
Time off:
12 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
ALTERNATE or SIMULTANEOUS
Waveform:
Asymetric Biphasic
Output:
0 mA
0 – 99 mA
NMES RE-ED:
Activation of muscle with electrical stimulation at a high intensity for a short time reduces disuse atrophy and
improves strength and power. The patient should participate to the extent possible by contracting during the
stimulation time.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Phase duration:
200 µS
30 µS – 500 µS
Pulse rate:
50 Hz
.0 Hz to 999 Hz
Phase duration mod:
Modulation time:
Modulation:
Auto Intensity:
0 %
0 – 20%
Treatment time:
15 minutes
0 – 99 minutes
Channel B delay:
0.0 seconds
.0 – 9.9 seconds
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
10 seconds
1 – 30 seconds
Time off:
50 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
B DELAYED, ALTERNATE or
SIMULTANEOUS
Waveform:
Asymetric Biphasic
Output:
0 mA
0 – 99 mA