Original Instructions
Cooling Fans
Cooling Fans
The Cooling Fans draw in ambient air from outside the oven and circulates it through the component compartments
to keep them within their operating temperature range
Testing
With Connector Plug disconnected check for resistance
BU to BR ≈ 200 Ω
BU to BK ≈ 695 Ω
BR to BK ≈ 900 Ω
Removal and Replacement
1 Disconnect oven from Power Supply
2 Remove Top and Right Side Panels (see Panel Removal)
3 Disconnect Harness Plug to Cooling Fans
4 Carefully lay oven on left side to access bottom
⚠
CAUTION
Oven is heavy.
5 Remove Phillips Head Screw securing Cooling Fan
Access Panel Disengage Panel and remove
6 Remove six (6) Screws securing Cooling Fan Mounting
Plate and pull from oven
7 Remove Screws (4 per fan) securing Cooling Fan to
Mounting Plate and remove Cooling Fan
8 To replace, reverse these steps
⚠
CAUTION
Bottom of Unit
Содержание ARX1
Страница 28: ... 2018 ACP Inc Cedar Rapids IA 52404 Page 28 of 80 Part No 16400033 Original Instructions Care and Cleaning 6 ...
Страница 77: ... 2018 ACP Inc Cedar Rapids IA 52404 Page 77 of 80 Part No 16400033 Original Instructions Notes ...
Страница 78: ... 2018 ACP Inc Cedar Rapids IA 52404 Page 78 of 80 Part No 16400033 Original Instructions Notes ...
Страница 79: ... 2018 ACP Inc Cedar Rapids IA 52404 Page 79 of 80 Part No 16400033 Original Instructions Notes ...
Страница 80: ... 2018 ACP Inc Cedar Rapids IA 52404 Page 80 of 80 Part No 16400033 Original Instructions www acpsolutions com ...