Waste Disposal
Check Bacteria Filter
The bacterial/hydrophobic filter is an important device to help protect the pump
and environment from moisture and bacteria taken from patient airways.
Clements recommend that the filter be changed every month or when damp or
discoloured.
Change after every use when the unit is used on infectious patients.
After Every 100 Hours or 2 Months of Operation
•
Check all suction tubing and replace if it is perished, soft or
discoloured. Also check the tubing connectors.
•
Check the seal ring on the lid and replace if hard, cracked or
perished. Check the fit of the lid in the collection jar (canister).
•
Check the overflow cut off valve seal and replace if perished or
damaged.
Waste Materials
The contents of the collection jars, suction tubing, bacteria filter,
internal exhaust filter may contain biohazardous wastes. Handle
using safe handling procedures, which may include the use of rubber
gloves and eye protection, and dispose of according to local protocols
for biohazardous materials.
Recycling
At the end of their service life, the pump and accessories should be
dismantled if necessary, and disposed of according to the WEEE
directive.