16
Troubleshooting
1.
The Pump Fails To Operate
•
Check that the mains power supply lead is firmly connected
and that the power is on at the mains power supply outlet.
•
Check fuse as detailed below.
•
Ensure that the Vacuum Gauge on the Control Panel indicates
zero. If not, release vacuum and switch pump off and on.
2.
Pump Running But No Vacuum
•
Check that all fittings are connected tightly.
•
Turn the vacuum controller knob on the control panel and
watch the indicator on the vacuum gauge.
•
Ensure that the silicone medical suction tubing is in good
condition and not old or cracked.
•
Check the type of handpiece in use as some handpieces
require finger occlusion or trigger release to allow suction.
•
Check bacteria/hydrophobic filter.
If the pump fails to work and you are unsure why, contact your
service department or return unit to your distributor.
Fuse Replacement
The ACeevac is fitted with a fuse located in a fuse holder that is
externally accessible
1. Disconnect the pump from the mains power supply.
2. Locate the fuse holder which is mounted on the rear side of
the unit at the lower right-hand side.
3. Pull out the fuse holder.
4. Remove the old fuse and inspect. If blown, replace it with a
new fuse of the same rating.
5. Replace the fuse holder cover.
Troubleshooting