12
Suction Accessories
Suction Accessories
The device and its accessories are biocompatible in accordance with
EN 60601-1.
Suction Cannulae
Suction cannulae or suction probes for contact with the human body
should comply with ISO 10993-1 requirements for biocompatibility.
Collection Jar (Canister)
The mechanical strength of the reusable collection jar (canister) is
guaranteed for 30 cycles of cleaning and sterilization. Beyond this,
the there may be signs of decay and replacement is recommended.
Silicone Tubing
The number of cleaning and sterilization cycles of the silicone tubing
is dependent on the actual usage. Tubing should be checked for
cracking and other visible signs of wear before re-use.
Disposable Liner Disconnection
1.
Switch off pump.
2.
Disconnect white elbow together with patient tubing from
“PATIENT” port and close port firmly with attached cap [8].
3.
Disconnect yellow nipple [6] from “VACUUM” port [5].
4.
Ensure that “TANDEM” port [3] is firmly closed with attached
cap [4].
5.
Carefully remove liner from support jar using lid handle and
dispose of according to local protocols for potentially
biohazardous waste.