CLEMENTS
CLEMENTS
CLEMENTS
CLEMENTS
5
Specifications
Specifications
Maximum Vacuum
-
80
kPa [
-
600 mmHg]
Maximum Flow
21 L/min (through jar) 32 L/min (free air)
Mains Power
Requirement
220
-
240
V~ 50 Hz 110 VA
Status Indicator
Green neon lamp in On/Off switch indicates power
available status
Pump
Piston type
Filter
Disposable hydrophobic and bacterial filter
BFE 99.9999%, VFE 0.027 micron
Collection Container
1 litre autoclavable, shatterproof plastic jar
Optional 2 litre autoclavable, shatterproof plastic jar
Optional 1 litre disposable liner jar
Optional 2 litre disposable liner jar
Overfill Protection
Float valve mechanism
Vacuum Control
Needle valve
Gauge
Bourdon tube type. Dual scale.
Gauge Range
0 to
-
100
kPa graduated at 5
kPa
0 to
-
760
mmHg graduated at 50
mmHg
Weight
3.6
kg
Dimensions
360L x 215W x 195H mm
Standard
Conditions
25°C, Sea level, 100kPa
Motor
PSC AC motor
Fuse
F1.6A L 250V
Jar Classifications
GMDN
44943
GMDN Term
Suction System Canister, Reusable
Device Class
Class I (Rule 1)
Sterilisation
Not supplied in sterile state