CLEMENTS
CLEMENTS
CLEMENTS
CLEMENTS
20
Periodic Safety Check
Periodic Safety Check
The following safety checks should be performed at least every 12
months by a qualified person who has adequate training, knowledge,
and practical experience to perform these tests.
* Inspect the equipment and accessories for mechanical and functional
damage.
* Inspect the relevant safety labels for legibility.
* Verify that the device functions properly as described in the
instructions for use.
* Perform electrical safety check.
Equipment Return for Repair
Before returning the ACeevac for repair, the external
surfaces and any accessories
must
be carefully
disinfected with a cloth soaked in methylated spirits or
hypochlorite-based
solution.
The
pump
and
any
accessories should then be placed in a bag with a note
outlining the disinfection undertaken.
Failure to follow
this procedure will result in the pump being returned
unrepaired
. Equipment returned for repair
must
be
accompanied by a description of the problem. Unless
specifically requested, return the pump and power cable
only, excluding accessories such as jar, filter and tubing.
IMPORTANT
There are no user-serviceable components inside.
Maintenance must be carried out by qualified
personnel only.