69
DO Setup:
Set Salinity / Set Alarm Limits
XL40 and 60 meters
Set Salinity
Water containing high levels of dissolved salts reduces dissolved oxygen in that water versus
water with little or no dissolved salts. The dissolved salts alter the relationship between the partial
pressure of oxygen and the oxygen concentration in water. By setting the salinity (in ppt) of your
sample, the XL40 and XL60 meter will compensate for the impact of salinity on the partial
pressure/oxygen concentration relationship of your sample.
The salinity can be set from 0 to 45 ppt.
To set Salinity
1. The current internal salinity value is displayed on the
Salinity box
.
2. Touch the
Salinity box
and use the numeric keypad on the screen to enter the salinity of
your sample. The
BS
key allows you to remove characters that are incorrectly entered.
3. Touch
Enter
on the keypad to accept the salinity value and return to the Setup screen.
Set Alarm Limits
This option allows you to set alarm limits for the DO measuring mode. If the DO value of the
measurement is outside of the boundaries set by the minimum and maximum limits, an audio and
visual warning will appear to let you know that your sample measurement was outside of the set
limits.
The Alarm Limit can be set from 0 to 600% Sat, 0 to 60 mg/L and 0 to 1272 mbar at the
atmospheric pressure of 1013 mbar.
To set Alarm Limits
1. Touch
ON
or
OFF
to set the status of the alarm of DO mode
2. Touch the
Low box
and use numeric keypad to enter the new limit values. The
BS
key
allows you to remove characters that are incorrectly entered.
3. Touch
Enter
on the keypad to accept this limit and return to the Setup screen.
4. Touch the
High box
and use numeric keypad to enter the new limit values. The
BS
key
allows you to remove characters that are incorrectly entered.
5. Touch
Enter
on the keypad to accept this limit and return to the Setup screen.
Содержание XL 20
Страница 1: ......
Страница 138: ...127 Ion Operation Known Addition Method XL25 50 and 60 meters...
Страница 140: ...129 Ion Operation Known Subtraction Method XL25 50 and 60 meters...
Страница 142: ...131 Ion Operation Analate Addition Method XL25 50 and 60 meters...
Страница 144: ...133 Ion Operation Analate Subtraction Method XL25 50 and 60 meters...