89
Conductivity Setup:
Data Storage Criteria
XL20, 30, 50 and 60 meters
This screen allows you to select which criteria are stored in the data logger with the measurement
when you store the data .The status of the current data storage criteria is displayed on the
screen. The criteria option is active if
ON
appears to the right of the option. It is inactive if
OFF
appears to the right of the option. Any active criteria will be stored on demand.
To set Data Storage Criteria
1. Touch the Touch here to edit button next to the Data Storage Criteria to access the
Conductivity Data Storage Criteria Setup screen.
2. Touch
ON
or
OFF
to change the status of a criteria you want to modify.
3. Repeat step
2
for all the remaining criteria except Data logging Interval (
Data storage
Interval
).
4. For Data logging Interval (
Data storage Interval
), touch
MANUAL
to log conductivity data
only when the Log Data button is pushed, touch
STABLE
to automatically log conductivity
data when conductivity reading is stable, or touch
TIMED
to set a specific timed interval in
seconds to log conductivity data.
5. Touch
OK
button to accept the changes of the entire group of data storage criteria and
return to the conductivity Setup screen.