DO Operation:
OUR Measurement
XL40 and 60 meters
This mode is used to measure Oxygen Uptake Rate (OUR). The OUR is used to determine the
oxygen consumption or respiration rate, is defined as the milligram of oxygen consumed per hour.
Remember to setup your OUR measuring mode parameters from the OUR Setup screen.
To Take OUR Measurements
1. Touch
OUR Start
icon to being OUR measurement.
2. The screen will indicate the test is in progress with
RUNNING
and a timer on the main
screen (the time the test is run is setup in the OUR Setup option, see page 38). The
OUR value in mg/L/hr and the DO reading appears as mg/L.
3. To end the OUR test, touch
Stop OUR
icon or let the timer expire.
If you touch the
Stop OUR
icon before the timer expires, you will get a message
indicating your measurement reading might not be accurate.
If you let the timer expire, the message states the measurement has been stopped
4. Touch
OK
to confirm either message.
5. Another dialog box appears asking if you wish to save current OUR result. Touch
OK
to save or
Cancel
to erase and return to OUR measurement screen.
6.
If you touched
OK
, a new dialog box appears to ask if you want to view the results.
Touch
OK
to proceed to the
Result View
screen showing OUR value and other
important test details.
7.
Touch
Exit
to exit Result View screen and return to OUR measurement screen. Touch
Delete All
to delete all OUR test results in memory. OR Touch
to send the OUR
test results to a computer or printer.
OUR measurement in progress
OUR Measurement View Result Screen
Содержание XL 20
Страница 1: ......
Страница 138: ...127 Ion Operation Known Addition Method XL25 50 and 60 meters...
Страница 140: ...129 Ion Operation Known Subtraction Method XL25 50 and 60 meters...
Страница 142: ...131 Ion Operation Analate Addition Method XL25 50 and 60 meters...
Страница 144: ...133 Ion Operation Analate Subtraction Method XL25 50 and 60 meters...