1.

 

Means of Delivery 

 

 

 

The bed is supplied on a transport bracket.  An 
inspection must take place upon receipt to ensure the 
delivery is complete and undamaged. 
 
Any missing parts, faults or damage must be reported 
immediately to the carrier, and Accora, in writing. 
 
When moving the bed, or loading and unloading it, 
care must be taken that the bed castors can rotate 
freely and the castor brakes are unlocked. The bed 
castors are designed for use in an indoor environment 
and for travel on even, smooth and clean floors, (e.g. 
ceramic floor tiles, linoleum, cast floors). The castors 
may become damaged when moving the bed along 
a rough, uneven or dirty surface. 

 

2.

 

Safety Instructions 

 

1.

 

Before using the bed, you must read the manual 
and use the bed in accordance with it. 

2.

 

The bed may not be used if faults have been 
detected on it that may injure the patient, staff or 
a third person, the bed or the surroundings. 

3.

 

The bed may only be operated by persons who 
are able to operate it in accordance with the 
manual. 

4.

 

The operating staff should acquaint the patient 
with the control functions that apply to the patient. 

5.

 

Before using the bed, the operator should 
understand the bed and its functionality. 

6.

 

The safe working load, as specified in section 4, 
must never be exceeded. 

7.

 

If there is a patient on the bed, the bed castors 
must be locked as an unlocked bed castor can 
cause injury to a patient who leaves the bed or 
changes position. 

8.

 

The height of the mattress platform must be 
adjusted to the correct height for the condition of 
the patient. 

9.

 

Only one person should occupy the bed at any 
time. 

10.

 

When operating the moving parts of the bed, care 
must be taken to ensure that the patient, other 
people and objects do not become trapped. 

11.

 

If a lifting pole or infusion stand is fixed to the bed, 
increased attention must be taken during 
movement, lifting or tipping, to the space around 
the lifting pole and infusion stand, so that the 
equipment is not damaged. 

12.

 

Before cleaning the bed, the electrical supply must 
be disconnected. 

13.

 

The bed may not be used where there is a danger 
of explosion or in the presence of inflammable 
liquids. 

14.

 

When repairing the bed, only original materials 
and components may be used, otherwise the 
manufacturer cannot guarantee against any 
damage that might occur. 

 

3.

 

Intended use 

 

This bed is intended for use in the following application 
environments only: 
 
1.

 

Application Environment 3 - Long-term care in a 
medical area where medical supervision is 
required and monitoring is provided if necessary 
and ME equipment used in medical procedures 
may be provided to help maintain or improve the 
condition of the patient. Note, this includes use in 
nursing homes and in rehabilitation and geriatric 
facilities. 

2.

 

Application Environment 4 - Care provided in a 
domestic area where ME equipment is used to 
alleviate or compensate for an injury, disability or 
disease 

 
It is essential to consult Accora in advance if you wish 
to use the bed for any purpose outside the use 
detailed in this manual. Electrical installations must 
meet local requirements. It is recommended that the 
bed is disconnected from the mains during 
exceptional cases (i.e. a storm). 

 

4.

 

Technical Specification 

 

Environmental information: 
 

Condition 

Temperature 

Range 

Relative 

Humidity 

Atmospheric 

pressure 

Operating 

+10°C to +40°C 

(+50°F to +104°F) 

30% to 75% 

(Non-

condensing) 

700 hPa to 

1060 hPa 

Transport/ 

storage 

-20°C to +50°C 

(-4°F to +122°F) 

 
If the bed is stored in conditions outside the normal 
operating range, it should be allowed time to stabilise, 
in normal operating conditions, before use. 

WARNING 

 
Extreme caution must be taken when moving 
the bed on the transport bracket to prevent 
the bed tipping over or moving unexpectedly. 

Содержание FloorBed 2

Страница 1: ......

Страница 2: ...is manual is important for realising the full potential of your new FloorBed please read it before use If you need further information please contact us at Tel 01223 206100 Fax 01223 206120 E mail inf...

Страница 3: ...electric hazards The bed must not be used if there is any visible damage to the handset or cable 13 Inappropriate use of the power supply cord e g kinking shearing could lead to dangerous electric ha...

Страница 4: ...trapped 11 If a lifting pole or infusion stand is fixed to the bed increased attention must be taken during movement lifting or tipping to the space around the lifting pole and infusion stand so that...

Страница 5: ...rking load is calculated as follows as specified by EN 60601 2 52 Maximum patient weight 150kg 23 5st 330lb Mattress 20kg 3st 44lb Accessories 15kg 2 5st 33lb TOTAL Safe working load 185kg 29st 407lb...

Страница 6: ...ress platform head section c Mattress platform foot section d Footboard e Headboard WARNING Assembly MUST be carried out by a competent person All functions MUST be tested and approved by a competent...

Страница 7: ...sert locking pin and R clip as shown in two positions 6 Insert two mattress platform joining bars and secure with locking pins and R clips as shown in two positions 7 Join the two parts of the bed tog...

Страница 8: ...oned If the medical staff state that the patient s medical condition is inappropriate for the patient to be able to adjust the bed independently the bed s position must be adjusted by the carer Please...

Страница 9: ...it stops in the Safety Stop position button 6 c Use the safety double button function to lower the bed to the Floor level position buttons 6 7 pressed together d Check all cables for risk of crushing...

Страница 10: ...s shown in Section 7 a Footboard b Headboard c Mattress platform foot section d Mattress platform head section 9 Ensure all R clips and pins are refitted so that the bed on the transport bracket is sa...

Страница 11: ...ean as described in the Cleaning Information 16 Troubleshooting The FloorBed does not function correctly 1 Is the Power on indicator light on the handset lit If not a Is the power supply plugged in an...

Страница 12: ...thin the scope of the Lifting Equipment Operations Regulations LOLER and so there is no formal legal requirement for regular inspection by a LOLER competent person The following checks must be carried...

Страница 13: ...dance RF emissions CISPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equi...

Страница 14: ...1kV line s to line s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and v...

Страница 15: ...d strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equi...

Страница 16: ...smitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2...

Страница 17: ...1 2010 We Accora Ltd Of Charter House Barrington Road Orwell Cambridge SG8 5QP hereby declare that Model reference FloorBed 2 Serial No Refer to product label conforms to the Requirements of Directive...

Страница 18: ...20 Accora 01223 206100 info accora uk com www accora uk com IFU FB2 001 EN Rev 1 JAN 2016...

Страница 19: ......