17. Storage 

 
For problem-free storage we recommend: 
 
1.

 

Disconnecting the bed from the electric supply 

2.

 

Remove the accessories  

3.

 

Wrapping the bed and accessories or covering 
them so that the coating and plastic parts are not 
damaged 

4.

 

Bed should be stored in a temperature between 
+10°C and +40°C 

5.

 

Bed should be stored somewhere where the 
relative humidity is between 30% and 75% 
 

18. Daily Inspection 

Daily visual inspection is strongly recommended and 
may be carried out by Carer, user or other (not 
necessarily a competent) person.  
 
The following checks must be carried out: 
 
1.  Does the bed operate as per its intended purpose 

without unexpected noise or motion? 

2.  Are there any signs of abuse or excessive wear? 
3.  Are all fixtures and fittings tight and secure? 
4.  Does the bed frame appear stable and secure? 
5.  Are all accessories fitted in line with the accessory 

manufacturer or accessory supplier’s instructions? 

6.  Are all the castors brakes in the locked position? 
7.  Are all electrical cables (including accessories, e.g. 

mattress air pump) secured and routed to prevent 
damage? 

8.  Does the handset locking function work correctly? 

(See Section 8) 

9.  Does the bed stop at the Safety Stop position? (See 

Section 9) 

10. Is the area around, above and below bed clear of 

possible obstruction? 

11. Is there any risk of entrapment or patient injury? 
 
If any damage, performance issue or cause for 
concern is noted during this inspection the bed should 
be withdrawn from service and appropriate steps 
should be taken. 

 
19. General maintenance

 

 

 

 
General maintenance must be carried out once yearly 
as a minimum and by a competent person or person 
trained to do so. The following checks must be carried 
out in addition to the checks listed in Section 20 – Daily 
Inspection. 
 
Note: It is not current practice for lifting beds to be 
considered to be lifting equipment within the scope of 
the Lifting Equipment Operations Regulations (LOLER), 
and so there is no formal legal requirement for regular 

inspection by a (LOLER) competent person. 
 
The following checks must be carried out: 
 
1. 

Do the castors operate correctly, including the 
brakes? 

2. 

Do all electrical bed positioning functions work 
correctly? (Backrest motion etc.). 

3. 

Does the power supply, or handset, show any sign 
of damage or abuse? 

4. 

Are any electrical cables pinched, crushed or 
damaged in any other way? 

5. 

Does the bed frame show any signs of abuse, 
damage or breakage? 

6. 

Is the bed frame mechanically sound without any 
cracking at welds etc? 

7. 

Is the control box plastic clip on cover holding the 
actuator plugs in place fitted? 

8. 

Are the plastic retaining covers fitted to the head 
and footboard actuators securing the wires in 
place? 

9. 

Is the control box metal cover fitted correctly? 

10.  Are all fixtures, fittings, nuts, bolts etc. tight and 

secure as appropriate? 

11.  Are all actuators secured using a 10mm clevis pin 

and R-clip? 

12.  Is the mattress platform secured to the head and 

footboards using 4 x 10mm clevis pin and R-clips? 

13.  Are the mattress platform joining bars clamped in 

place (with M8 bolts from below) and secured 
with 4 x 10mm clevis pin and R-clips? 

 
If any damage, performance issue or cause for 
concern is noted during this inspection the bed should 
be withdrawn from service and appropriate steps 
should be taken. 
 

20. Guarantee 

 

The warranty period for the FloorBed is 24 months, 
covering parts only.  
 
If the bed was acquired for a care home, the care 
home maintenance department should contact 
Accora to obtain the necessary parts. 
 

21. Disposal of the FloorBed 

 

In the event of the disposal of materials from the bed, 
end-of-life parts must be disposed of in accordance 
with current environmental regulations. 

WARNING 

 

Maintenance MUST be carried out by a 
competent person.  
All functions MUST be tested and approved by 
a competent person after maintenance. 
Refer to Sections 8 and 9. 
Only power supply supplied with bed should 
be used. 
Do not carry out maintenance with service 
user or patient on the bed.  
 

Содержание FloorBed 2

Страница 1: ......

Страница 2: ...is manual is important for realising the full potential of your new FloorBed please read it before use If you need further information please contact us at Tel 01223 206100 Fax 01223 206120 E mail inf...

Страница 3: ...electric hazards The bed must not be used if there is any visible damage to the handset or cable 13 Inappropriate use of the power supply cord e g kinking shearing could lead to dangerous electric ha...

Страница 4: ...trapped 11 If a lifting pole or infusion stand is fixed to the bed increased attention must be taken during movement lifting or tipping to the space around the lifting pole and infusion stand so that...

Страница 5: ...rking load is calculated as follows as specified by EN 60601 2 52 Maximum patient weight 150kg 23 5st 330lb Mattress 20kg 3st 44lb Accessories 15kg 2 5st 33lb TOTAL Safe working load 185kg 29st 407lb...

Страница 6: ...ress platform head section c Mattress platform foot section d Footboard e Headboard WARNING Assembly MUST be carried out by a competent person All functions MUST be tested and approved by a competent...

Страница 7: ...sert locking pin and R clip as shown in two positions 6 Insert two mattress platform joining bars and secure with locking pins and R clips as shown in two positions 7 Join the two parts of the bed tog...

Страница 8: ...oned If the medical staff state that the patient s medical condition is inappropriate for the patient to be able to adjust the bed independently the bed s position must be adjusted by the carer Please...

Страница 9: ...it stops in the Safety Stop position button 6 c Use the safety double button function to lower the bed to the Floor level position buttons 6 7 pressed together d Check all cables for risk of crushing...

Страница 10: ...s shown in Section 7 a Footboard b Headboard c Mattress platform foot section d Mattress platform head section 9 Ensure all R clips and pins are refitted so that the bed on the transport bracket is sa...

Страница 11: ...ean as described in the Cleaning Information 16 Troubleshooting The FloorBed does not function correctly 1 Is the Power on indicator light on the handset lit If not a Is the power supply plugged in an...

Страница 12: ...thin the scope of the Lifting Equipment Operations Regulations LOLER and so there is no formal legal requirement for regular inspection by a LOLER competent person The following checks must be carried...

Страница 13: ...dance RF emissions CISPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equi...

Страница 14: ...1kV line s to line s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and v...

Страница 15: ...d strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equi...

Страница 16: ...smitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2...

Страница 17: ...1 2010 We Accora Ltd Of Charter House Barrington Road Orwell Cambridge SG8 5QP hereby declare that Model reference FloorBed 2 Serial No Refer to product label conforms to the Requirements of Directive...

Страница 18: ...20 Accora 01223 206100 info accora uk com www accora uk com IFU FB2 001 EN Rev 1 JAN 2016...

Страница 19: ......

Отзывы: