14.

 

Moving and Repositioning 

 

 

 

1.

 

Ensure the bed is at the Safety Stop position (see 
Section 9, point b). 

2.

 

Disconnect the power. 

3.

 

Secure the handset, power supply and all cables to 
prevent damage. 

4.

 

Unlock the castors and move the bed. 

5.

 

When the bed has been moved or repositioned 
lock al the castors and follow the instruction in 
Section 8 

6.

 

When the bed has been positioned correctly, follow 
the instructions in Section 9. 

 

15.

 

Cleaning & Disinfection 

 

 

 

Cleaning Information: 

 
In order to disinfect the bed, only use detergents 
designed for use in healthcare. Do not use abrasives, 
scourers or other materials that could damage the 
coating. Do not use corrosives, caustics or strong acids. 
Do not use detergents that could alter the structure or 
behaviour of the plastics (petrol etc.). 

 
Clean by wiping with a damp and properly wrung 
cloth. 
 
The bed is not designed for maintenance in automatic 
bed washers or for cleaning with pressurised water, 
spraying, showering or steam cleaning. 
 
Accora cannot be liable for any damage or risk of 
damage if inappropriate cleaning or disinfectant 
agents are used. 

 

Cleaning procedure: 

 
1.

 

Remove all accessories, mattress etc. 

2.

 

Adjust the mattress platform to the highest position 
and adjust the position of the Backrest and Footrest 
to provide access for cleaning all the platform 
parts. 

3.

 

Disconnect the bed from the power supply. 

4.

 

Move the bed to where cleaning will take place 
and lock the bed brakes. 

5.

 

Clean as described in the “Cleaning Information”. 

 

16. Troubleshooting 

 

 

 
The FloorBed does not function correctly: 
 
1.

 

Is the ‘Power on’ indicator light on the handset lit? If 
not: 
a.

 

Is the power supply plugged in and turned on? 

b.

 

Is the power lead in-line socket plugged in 
correctly? 

c.

 

If the ‘Power on’ indicator light is still not lit, 
contact Accora for further advice. 

2.

 

If the ‘Power on’ indicator light is on, has the 
handset been unlocked? If not, refer to Section 8. 

3.

 

If the handset has been unlocked and the FloorBed 
still does not function correctly, contact Accora for 
further advice. 

 
If the FloorBed does not stop at the Safety-stop 
position, check the main lift actuators are correctly 
plugged in at all points between the actuator and the 
control box. 
 
If the FloorBed still does not function correctly, contact 
Accora for further advice. 
 

WARNING 

 

Moving or repositioning MUST be carried out 
by a competent person.  
All functions MUST be tested and approved by 
a competent person after moving or 
repositioning 
Only power supply supplied with bed should 
be used. 
Do not move the bed in the Floor-Level 
position 
Do not move or reposition the bed with 
service user or patient on the bed.  

WARNING 

 

Cleaning and disinfection MUST be carried out 
by a competent person.  
The bed must be disconnected from the 
power supply when being cleaned or 
disinfected. 
All functions MUST be tested and approved by 
a competent person after cleaning or 
disinfection. 

WARNING 

 

Troubleshooting MUST be carried out by a 
competent person.  
Do not attempt to open any electrical part 
enclosures. 
Do not attempt to repair any electrical parts. 
All functions MUST be tested and approved by 
a competent person after troubleshooting 
 
 

Содержание FloorBed 2

Страница 1: ......

Страница 2: ...is manual is important for realising the full potential of your new FloorBed please read it before use If you need further information please contact us at Tel 01223 206100 Fax 01223 206120 E mail inf...

Страница 3: ...electric hazards The bed must not be used if there is any visible damage to the handset or cable 13 Inappropriate use of the power supply cord e g kinking shearing could lead to dangerous electric ha...

Страница 4: ...trapped 11 If a lifting pole or infusion stand is fixed to the bed increased attention must be taken during movement lifting or tipping to the space around the lifting pole and infusion stand so that...

Страница 5: ...rking load is calculated as follows as specified by EN 60601 2 52 Maximum patient weight 150kg 23 5st 330lb Mattress 20kg 3st 44lb Accessories 15kg 2 5st 33lb TOTAL Safe working load 185kg 29st 407lb...

Страница 6: ...ress platform head section c Mattress platform foot section d Footboard e Headboard WARNING Assembly MUST be carried out by a competent person All functions MUST be tested and approved by a competent...

Страница 7: ...sert locking pin and R clip as shown in two positions 6 Insert two mattress platform joining bars and secure with locking pins and R clips as shown in two positions 7 Join the two parts of the bed tog...

Страница 8: ...oned If the medical staff state that the patient s medical condition is inappropriate for the patient to be able to adjust the bed independently the bed s position must be adjusted by the carer Please...

Страница 9: ...it stops in the Safety Stop position button 6 c Use the safety double button function to lower the bed to the Floor level position buttons 6 7 pressed together d Check all cables for risk of crushing...

Страница 10: ...s shown in Section 7 a Footboard b Headboard c Mattress platform foot section d Mattress platform head section 9 Ensure all R clips and pins are refitted so that the bed on the transport bracket is sa...

Страница 11: ...ean as described in the Cleaning Information 16 Troubleshooting The FloorBed does not function correctly 1 Is the Power on indicator light on the handset lit If not a Is the power supply plugged in an...

Страница 12: ...thin the scope of the Lifting Equipment Operations Regulations LOLER and so there is no formal legal requirement for regular inspection by a LOLER competent person The following checks must be carried...

Страница 13: ...dance RF emissions CISPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equi...

Страница 14: ...1kV line s to line s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and v...

Страница 15: ...d strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equi...

Страница 16: ...smitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2...

Страница 17: ...1 2010 We Accora Ltd Of Charter House Barrington Road Orwell Cambridge SG8 5QP hereby declare that Model reference FloorBed 2 Serial No Refer to product label conforms to the Requirements of Directive...

Страница 18: ...20 Accora 01223 206100 info accora uk com www accora uk com IFU FB2 001 EN Rev 1 JAN 2016...

Страница 19: ......

Отзывы: