Guidance and manufacturer’s declaration-electromagnetic immunity
The bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the bed should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-
guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply
lines
+ 1kV for
input/output
lines
+ 2kV for power
supply
lines
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge IEC 61000-4-5
+ 1kV line(s) to line(s)
+ 2kV line(s) to earth
+ 1kV differential
mode
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips, short
interruptions and
voltage
variations on power
supply
input lines IEC 61000-4-
11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the bed
requires continued
operation during
power mains
interruptions, it is
recommended that
the bed be powered
from an uninterruptible
power supply or a
battery.
Power frequency(50,
60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The bed power
frequency magnetic
fields should be at
levels characteristic of
a typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.