Guidance and manufacturer’s declaration-electromagnetic immunity 

The bed is intended for use in the electromagnetic environment specified below. 

The customer or the user of the bed should assure that it is used in such an environment. 

Immunity test 

IEC 60601 
test level 

Compliance level 

Electromagnetic 
environment-
guidance 

Electrostatic 
discharge(ESD) 

IEC 61000-4-2 

+ 6 kV contact 

+ 8 kV air 

+ 6 kV contact 

+ 8 kV air 

Floors should be wood, 
concrete or ceramic 
tile. If floors are 
covered with synthetic 
material, the relative 
humidity should be at 
least 30% 

Electrical fast 
transient/burst 

IEC 61000-4-4 

+ 2kV for power 
supply 

lines 

+ 1kV for 
input/output 

lines 

+ 2kV for power 
supply 

lines 

Not applicable 

Mains power quality 
should be that of a 
typical commercial or 
hospital environment. 

Surge IEC 61000-4-5 

+ 1kV line(s) to line(s) 

+ 2kV line(s) to earth 

+ 1kV differential 
mode 

Not applicable 

Mains power quality 
should be that of a 
typical commercial or 
hospital environment. 

Voltage Dips, short 

interruptions and 
voltage 

variations on power 
supply 

input lines IEC 61000-4-
11 

<5% UT(>95% dip in 

UT) for 0,5 cycle 

40% UT(60% dip in 

UT) for 5 cycles 

70% UT(30% dip in 

UT) for 25 cycles 

<5% UT(>95% dip in 

UT) for 5 s 

<5% UT(>95% dip in 

UT) for 0,5 cycle 

40% UT(60% dip in 

UT) for 5 cycles 

70% UT(30% dip in 

UT) for 25 cycles 

<5% UT(>95% dip in 

UT) for 5 s 

Mains power quality 
should be that of a 
typical commercial or 
hospital environment. 
If the user of the bed 
requires continued 
operation during 
power mains 
interruptions, it is 
recommended that 
the bed be powered 
from an uninterruptible 
power supply or a 
battery. 

 

Power frequency(50, 
60 Hz) 

magnetic field IEC 
61000-4-8 

3 A/m 

3 A/m 

The bed power 
frequency magnetic 
fields should be at 
levels characteristic of 
a typical location in a 
typical commercial or 
hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

Содержание FloorBed 2

Страница 1: ......

Страница 2: ...is manual is important for realising the full potential of your new FloorBed please read it before use If you need further information please contact us at Tel 01223 206100 Fax 01223 206120 E mail inf...

Страница 3: ...electric hazards The bed must not be used if there is any visible damage to the handset or cable 13 Inappropriate use of the power supply cord e g kinking shearing could lead to dangerous electric ha...

Страница 4: ...trapped 11 If a lifting pole or infusion stand is fixed to the bed increased attention must be taken during movement lifting or tipping to the space around the lifting pole and infusion stand so that...

Страница 5: ...rking load is calculated as follows as specified by EN 60601 2 52 Maximum patient weight 150kg 23 5st 330lb Mattress 20kg 3st 44lb Accessories 15kg 2 5st 33lb TOTAL Safe working load 185kg 29st 407lb...

Страница 6: ...ress platform head section c Mattress platform foot section d Footboard e Headboard WARNING Assembly MUST be carried out by a competent person All functions MUST be tested and approved by a competent...

Страница 7: ...sert locking pin and R clip as shown in two positions 6 Insert two mattress platform joining bars and secure with locking pins and R clips as shown in two positions 7 Join the two parts of the bed tog...

Страница 8: ...oned If the medical staff state that the patient s medical condition is inappropriate for the patient to be able to adjust the bed independently the bed s position must be adjusted by the carer Please...

Страница 9: ...it stops in the Safety Stop position button 6 c Use the safety double button function to lower the bed to the Floor level position buttons 6 7 pressed together d Check all cables for risk of crushing...

Страница 10: ...s shown in Section 7 a Footboard b Headboard c Mattress platform foot section d Mattress platform head section 9 Ensure all R clips and pins are refitted so that the bed on the transport bracket is sa...

Страница 11: ...ean as described in the Cleaning Information 16 Troubleshooting The FloorBed does not function correctly 1 Is the Power on indicator light on the handset lit If not a Is the power supply plugged in an...

Страница 12: ...thin the scope of the Lifting Equipment Operations Regulations LOLER and so there is no formal legal requirement for regular inspection by a LOLER competent person The following checks must be carried...

Страница 13: ...dance RF emissions CISPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equi...

Страница 14: ...1kV line s to line s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and v...

Страница 15: ...d strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equi...

Страница 16: ...smitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2...

Страница 17: ...1 2010 We Accora Ltd Of Charter House Barrington Road Orwell Cambridge SG8 5QP hereby declare that Model reference FloorBed 2 Serial No Refer to product label conforms to the Requirements of Directive...

Страница 18: ...20 Accora 01223 206100 info accora uk com www accora uk com IFU FB2 001 EN Rev 1 JAN 2016...

Страница 19: ......

Отзывы: