ABEM Terraloc Pro 2
68
Linearity
range [dB]
-
This value sets the amplitude level that is within the linear part of the
scaling function. Both scaling functions are linear in the beginning (for
small amplitudes), while compressing larger amplitudes.
Only available when
View Mode
is set to Hyperbolic
Number of
traced to
display
-
This value sets the number of traces that will be displayed in the
Record View.
Only available if acquisition mode is set to Common Offset
Windowing
function
-
Selects the function to be used for data windowing.
Available values are:
No window, Hanning, Hamming, Blackman, Bartlett, Kaiser, 4th order
Blackman-Harris, Flat top
Max
frequency
[Hz]
-
The maximum frequency to display. The displayed spectrum will go
from 0 Hz up to the selected maximum frequency.
Available values are:
50, 100, 200, 500, 1000, 2000, 5000, 10000, 25000
Dynamic
range [dB]
-
The maximum frequency component is used as reference when
calculating the spectrum. The displayed spectrum will go from 0 dB up
to the selected maximum dB value.
Available values are: from 6 dB up to 198 dB in steps of 6 dB
4.6.14 The Velocity Analyzer
The
Velocity Analyzer
consists of a dialog, which is displayed at the top of the screen,
and the
Velocity Marker
, a red line that is displayed in the
Record View
The
Velocity Analyzer
can be used to estimate the apparent seismic velocity in
refraction records.