ABEM Terraloc Pro 2
55
120
150
132
165
144
180
156
195
168
210
180
225
192
240
Remove
calculated DC
offset after
measurement
-
The measurements are calibrated for internal offset, but offset
during measurements may appear depending on attached sensors
and site conditions. This option will calculate and remove any
DC offset from the measured data.
Table 12 Filter settings
4.6.6 The Noise Monitor Dialog
There is a real-time noise monitor integrated in the system. It is displayed in the
Noise
Monitor
dialog (Figure 47). The
Noise Monitor
can be used to just inspect the noise
level, or monitor the noise so that the operator is able to fire the shot at the right moment.
•
Opening the
Noise Monitor
dialog
―
The
Noise Monitor
is opened either directly from
Noise Monitor Setup
dialog
with the
Show Noise
button or when the
Noise Monitor Status
is turned on and
the instrument is armed
Figure 47 The Noise Monitor Dialog
•
Adjusting the graph min (Figure 48)
―
Press <SHIFT> + <+> to increase the graph min in 10 dB step
―
Press <SHIFT> + <-> to decrease the graph min in 10 dB step