ABEM Terraloc Pro 2
52
Ext. arm
verify
-
When several instruments are interconnected, the external arm
inputs and outputs can be connected in such a way that when one
instrument is armed it in turn will arm the next instrument. If this
choice is checked when the user presses <ARM> on one instrument,
it will wait until it receives an external arm from the last instrument
in the chain before actually accepting the arm event. If no external
arm is received within the timeout set, the instrument will disarm
and the disarm event will propagate to all the other instruments
Verify
timeout
[ms]
-
The time to wait for an external arm before disarming and showing
an error message
Table 10 External Arm/Trig settings
4.6.5.3
The Noise Monitor Settings Category
The
Noise Monitor Setup
dialog (Figure 45) has settings that control the
Noise Monitor
dialog (chapter 4.6.6).
•
Opening the
Noise Monitor Setup
dialog (
Acquisition Setup
dialog with the Noise
tab selected)
―
Press <3>
Figure 45 The Acquisition Setup Dialog; Noise tab selected
Setting
Description
Noise monitor
status
-
When the noise monitor is On, it will be displayed when the
instrument is armed. Available choices are: On, Off
Содержание Terraloc Pro 2
Страница 1: ...User Manual Terraloc Pro 2 ABEM 20181009 based on release 3 0 4 of SeisTW ...
Страница 6: ...ABEM Terraloc Pro 2 iv ...
Страница 73: ...ABEM Terraloc Pro 2 71 Free end to the left Free end to the right Free end up Free end down Free end left ...
Страница 74: ...ABEM Terraloc Pro 2 72 Free end up Moved to the right ...