Table 6. Possible causes and solutions for potential
issues
Problem
Possible Cause
Possible Solution
Turn on Bluetooth®
wireless technology using
the instructions in the
“Troubleshooting
messages” table.
Patient controller is
damaged or
malfunctioning.
Replace the patient
controller.
Cannot find or
connect to the
generator.
The device is not
supported.
Download the patient
controller app for your
system.
Use the device provided
by Abbott Medical or a
compatible personal
Apple‡ iOS‡ device.
NOTE: For more
information on
compatible devices, see
(page 73).
Generator
Generator
Unavailable
message appears.
The generator is not in
range.
Decrease the distance
between the generator
and patient controller.
Low Generator
Battery message
appears.
The EPG battery needs
to be replaced soon.
Contact your clinician to
schedule a replacement.
Generator Battery
Depleted message
appears.
The EPG battery level is
too low to provide
therapy.
Contact your clinician to
schedule a replacement.