Consumer goods and electronic devices.
Magnetic
interference with consumer goods or electronic devices
that contain magnets, such as mobile phones and smart
watches, may unintentionally cause the
neurostimulation system to turn on or turn off or affect
communication between the patient controller and
generator; however, it will not change the prescribed
programmed parameters. To prevent accidental
interference, keep your smart device at least 15 cm
(6 in.) away from the generator, and do not carry any
smart device in a pocket near the generator. Contact
Technical Support or your physician to discuss modifying
the magnet mode setting on your neurostimulation
system or if you have any other questions.
Additional Precautions Only Applicable to
the Trial System
Battery damage.
If the internal batteries are shorted, it
is possible for the external temperature of the pouch to
reach 42°C (107°F), which could damage the EPG. If this
occurs, contact your clinician.
Additional Precautions Only Applicable to
the Permanent System
High-output ultrasonics and lithotripsy.
The use of high-
output devices, such as an electrohydraulic lithotriptor,
may cause damage to the electronic circuitry of an
implanted device. If lithotripsy must be used, do not
focus the energy near the device.
Transcranial magnetic stimulation (TMS) and
electroconvulsive therapy (ECT).
Safety has not been
19