WaterMaster
Electromagnetic flowmeter
2 Parameter Tables
56
IM/WMPBST–EN Rev. B
41 / 25
SIMULATION_VOLUME_FLOW_RANGE_UPPER
Upper limit for internal simulated volume flowrate values.
Float
4
D
r / C / a
0
42 / 26
SIMULATION_FLOW_VELOCITY_RANGE_LOWER
Lower limit for internal simulated velocity values.
Float
4
D
r / C / a
0
43 / 27
SIMULATION_FLOW_VELOCITY_RANGE_UPPER
Upper limit for internal simulated velocity values.
Float
4
D
r / C / a
0
44 / 28
LOOP_RESISTANCE_RANGE_LOWER
Lower limit of loop resistance range.
Unit: Ohm
Float
4
D
r / C / a
15
45 / 29
LOOP_RESISTANCE_RANGE_UPPER
Upper limit of loop resistance range.
Unit: Ohm
Float
4
D
r / C / a
65
46 / 30
TRANSMITTER_AVERAGE_GAIN_SHIFT_RANGE_LOWER
Lower limit of transmitter average gain shift range.
Unit: %
Float
4
D
r / C / a
–0.4
47 / 31
TRANSMITTER_AVERAGE_GAIN_SHIFT_RANGE_UPPER
Upper limit of transmitter average gain shift range.
Unit: %
Float
4
D
r / C / a
0.4
48 / 32
DIAGNOSTIC_VALUE_UNITS
Units of the value currently being read by the Diagnostic Signal function block.
1281: Ohms
1284: kOhms
1240: V
1271: mH
1342: %
Unsigned 16
2
D
r / C / a
Ohms
49 / 33
ELECTRODE_1_RESISTANCE
Resistance of electrode 1.
Unit: kOhm
101
5
D
r / O / a
0
50 / 34
UNUSED
Unused parameter.
–
–
–
–
–
51 / 35
UNUSED
Unused parameter.
–
–
–
–
–
52 / 36
UNUSED
Unused parameter.
–
–
–
–
–
53 / 37
UNUSED
Unused parameter.
–
–
–
–
–
Abs. Idx /
Rel. Idx
Variable Name / Brief Description
Data Type
Size
Store
Access /
Usage /
Transport
Default
Table 2.14 Slot 12 – Transducer Block 5 – Diagnostics (Continued)