WaterMaster
Electromagnetic flowmeter
2 Parameter Tables
36
IM/WMPBST–EN Rev. B
95 / 79
LOW_FLOW_CUTOFF_HYSTERESIS
Hysteresis for flow cutoff in percent.
Min: 0
Max: 50
Unit: %
Float
4
S
r,w / C / a
20
96 / 80
METER_MODE
Flow direction under which metering takes place.
0: Normal Operation
1: Forward Flow
2: Reverse Flow
Unsigned 8
1
S
r,w / C / a
Normal Operation
97 / 81
UNUSED
Unused parameter.
–
–
–
–
–
98 / 82
UNUSED
Unused parameter.
–
–
–
–
–
99 / 83
PROBE_PIPE_BORE
Bore of pipe where probe type sensor is installed.
Min: 1
Max: 5000
Unit: mm
Float
4
S
r,w / C / a
250
100 / 84
PROBE_INSERTION_FACTOR
Insertion factor for probe type sensors.
Min: 0
Max: 3
Float
4
S
r,w / C / a
1
101 / 85
PROBE_PROFILE_FACTOR
Profile factor probe type sensors.
Min: 0
Max: 3
Float
4
S
r,w / C / a
1
102 / 86
Q_MAX_DN_SPECIAL
QMax DN value for special sensors.
Min: 0
Max: 100000
Unit: TB1.VOLUME_FLOW_UNITS
Float
4
S
r,w / C / a
2.5
103 / 87
USER_SYSTEM_SPAN
User span value.
Min: -250
Max: 250
Unit: %
Float
4
S
r,w / C / a
100
104 / 88
MAINS_FREQUENCY
Transmitter mains frequency.
Changes to this parameter cause the instrument to reboot.
50: 50
60: 60
Unit: Hz
Unsigned 8
1
S
r,w / C / a
50
105 / 89
FACTORY_SENSOR_SPAN
Factory-set sensor span value.
Unit: %
Float
4
D
r / C / a
100
Abs. Idx /
Rel. Idx
Variable Name / Brief Description
Data Type
Size
Store
Access /
Usage /
Transport
Default
Table 2.10 Slot 8 – Transducer Block 1 – Flow (Continued)