![ABB Navigator 550 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/abb/navigator-550/navigator-550_operating-instructions-manual_278033.webp)
Navigator 500
Sodium analyzer
6 Operation overview – transmitter
OI/ASO550–EN Rev. G
31
6.4 Navigation overview
6.5 Operating mode
In operating mode, values from connected wet-sections are displayed on
Operator
Pages
. A maximum of 5
Operator
Pages
can be
displayed.
Operator
Page
1
(the default page) displays values from all connected wet-sections simultaneously (a maximum of 4 wet-sections
can be connected). The remaining
Operator Pages
can be assigned to display values from individual wet-sections (in any wet-section
order). To achieve this, each wet-section must be associated with a template in the
Configuration
level /
Display
/
Operator Templates
parameters – see page 50.
Operator Page 1
shows that 4 wet-sections are connected.
Fig. 6.3 Menu navigation
*The highest priority diagnostic or alarm is displayed. Other active diagnostic / alarm states can be viewed on the
Diagnostics
View
– see page 33.
Fig. 6.4 Operator Page 1 (sodium multiple wet-section)
Operator
Page 1
Operator
Page 2
Operator
Page 3
Operator
Page 4
Operator
Page 5
Diagnostics
View
Signals
View
Chart
View
Alarms
View
Outputs
View
Alarm
Log
Audit
Log
Calibration
Log
Diagnostic
Log
2.77
2.77
4.28
4.28
6.93
6.93
No UNa
8.82
8.82
2012–04–15
10 : 29: 17
CAL
Instrument tag
(user-assignable)
Time and date
Display icon –
see Section 6.10, page 40
Unit of measure
NAMUR (NE107) Status Icons
Failure
High Process Alarm
Maintenance
Low Process Alarm
Out of Spec
High Latch Alarm
Check Function
Low Latch Alarm
Display icon –
Diagnostic message or alarm tag –
see Appendix A, page 72 for diagnostic messages
Wet-section tag
(user-assignable)
Process value (PV)
Parameter ID
Autoscroll
icon –
scrolls each page
sequentially when enabled
TAG 1
TAG 2
In Configuration
Navigator
TAG 3
ppb
ppb
ppb
Na
Na
Na
Na
ppb