Appendix A – Troubleshooting
OI/ASO550–EN Rev. G
73
NV Error SW Key 1
NV error – software key 1 (CRC Comms.)
Failure of non-volatile memory on software key 1 board or permanent corruption of its data.
Cycle power to the device.
If problem persists check all configuration parameters and correct any errors.
If problem still persists contact local service organization.
Temp Failure
(S1, S2, S3, S4)
Temperature sensor failure for wet-section1 (2, 3, 4).
The temperature compensator or associated connections are either open-circuit or short-circuit.
Check wiring temperature compensator connections to the PCB.
Calibrating
(S1, S2, S3, S4)
Calibrating.
Displayed during calibration of wet-section (1, 2, 3, 4).
On a multiple wet-section setup, this inhibits other wet-sections being calibrated.
In Hold Mode
(S1, S2, S3, S4)
Wet-section (1, 2, 3, 4) in manual hold mode via front panel.
Analog outputs and alarms are held.
To exit manual hold press the
key, scroll to
Manual Hold
and select the appropriate wet-section(s).
Recovery
(S1, S2, S3, S4)
Wet-section(s) performing a recovery stage after calibration, regeneration, or after exiting
Man.Valve Control
–
see page 47.
During the recovery period, outputs and alarms are held if
Hold Outputs
is enabled – see page 45.
Regeneration
A regeneration routine is being run.
Refer to Section 5.7, 28 for details of this routine.
Simulation On
The analyzer is operating in
Simulation
mode.
Cal.Failed
(S1, S2, S3, S4)
Last wet-section calibration failed.
Check reagent is entraining (bubbles at end of entrainment tube, check reagent is fresh).
Check the calibration solution.
Flow Error
(S1, S2, S3, S4)
Displayed only if
flowmeter is fitted
Sample flow rate is less than 50 mlmin.(3.05 cu in.min.)
Increase the sample flow to the wet-section.
Media Card Full
Memory card is full, no more data can be saved to the card.
Replace memory card.
Missed Cal.
(S1, S2, S3, S4)
Missed last schedule calibration.
PV Range
(S1, S2, S3, S4)
Process value (PV) measured is out of the specified range of the wet-section.
0 to 10,000 ppb.
Sample Cold
(S1, S2, S3, S4)
Sample solution temperature lower than 5 °C (41 °F).
Increase the temperature of the sample.
Sample Hot
(S1, S2, S3, S4)
Sample solution temperature higher than 55 °C (131 °F).
Reduce the temperature of the sample.
Icon
Diagnostic message
Possible cause and suggested action