Navigator 500
Sodium analyzer
7 Menu descriptions
OI/ASO550–EN Rev. G
45
Menu
Comment
Defaults
Calibration Setup
Low Std Solution
Sets the low standard calibration solution value between 50 and
10,000 ppb.
100 ppb
High Std Solution
Sets the high standard calibration solution value between 50 and
10,000 ppb.
1000 ppb
Recovery Time
Sets the recovery time in minutes (between 10 and 60 minutes).
This value can be adjusted to ensure the wet-section has adequate
time to reach a stable reading after a calibration.
The default setting is 15 minutes, but this may need to be extended if
the process concentration is very low; especially if a high calibration
has been carried out.
15 min.
Grab Sample Time
Sets the grab sampling time in minutes.
15 min.
Restore Defaults
Restores
Calibrate
parameters to their default values / settings.
Hold Outputs
Enables / Disables the
Hold Outputs
function.
Outputs include current outputs and alarms.