![ABB Navigator 550 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/abb/navigator-550/navigator-550_operating-instructions-manual_278039.webp)
Navigator 500
Sodium analyzer
6 Operation overview – transmitter
OI/ASO550–EN Rev. G
37
6.8.2 SD card / USB stick insertion and removal
To access the SD card / USB stick:
1. Ensure the transmitter is offline.
2. Use a large Pozi-drive screwdriver to release the door
catch
A
.
3. Open the door and insert the SD card / USB stick
B
.
The red LED
C
is lit when the SD card / USB stick is in
use by the transmitter.
4. To remove the SD card / USB stick, if the red LED is lit,
press button
D
and wait until the LED goes out.
5. Pull the SD card / USB stick out of its socket. The SD card
/ USB stick can then be inserted into an appropriate card
reader USB port on the PC and the data downloaded.
6.8.3 Archive file types
All archive files created by the transmitter (except for
configuration files) are assigned filenames automatically. Each
type of archived file is assigned a different file extension.
Archive files are created as text format data files.
The file type and extension for
Data
text files is '.DOO'
<ddmmyy><hhmmss><monitor tag>.DOO
The file type and extension for
Event
log files (containing
historical entries from the
Audit, Calibration, Diagnostic
and
Alarm
logs is '.AOO’.
<ddmmyy><hhmmss><monitor tag>.AOO
Configuration filenames are user-entered. The configuration file
type and extension is '.cfg’.
6.8.4 Data files
Text format archived data is stored in a comma-separated value
(CSV) format and can be imported directly into a standard
spreadsheet, for example, Microsoft
®
Excel.
Alternatively, you can carry out detailed graphical analysis of the
data on a PC using ABB’s DataManager data analysis software.
New data files are created in the following circumstances:
The transmitter configuration is changed.
One of the current files exceeds the maximum permissible
size (a new file is created at 12:00 a.m. on the following
day. Data is logged into the existing file continuously until
the new file is created.
When the daylight saving period starts or ends.
When working files cannot be found / are corrupted.
The filename is formatted as follows:
Data logs: <ddmmyy><hhmmss><monitor tag>.DOO
Fig. 6.12 SD card / USB stick insertion and removal
Note.
Data stored in the internal memory buffer can still be
transferred to the SD card / USB stick when the archive
media is placed on-line again (providing it is not off-line so
long that the un-archived data in the internal memory is
overwritten).
Note.
The 'monitor tag' is set in the
Device Setup
level (see
page 42) when the user has access at
Advanced
level
– see Section 6.9, page 39.
The time and date are formatted according to the
format set in
Display
/
Date & Time
level – see
page 42.
The transmitter's internal clock can be configured to
adjust automatically at the start and end of
Daylight
Saving