Navigator 500
Sodium analyzer
3 Installation
OI/ASO550–EN Rev. G
11
3.1.5 Connecting the reagent / calibration / regeneration tubing
The following tubing connections must be made on site, all other tubing connections are factory-made. All other internal wet-section
tubing connections are factory-made.
Referring to Fig. 3.5:
1. Connect the end of reagent tubing
A
to the connector
B
on the reagent bottle.
2. Connect the calibration solution 2 tubing QD coupling plug
C
to the mating connector
D
at the base of the calibration
solution 2 bottle.
3. If an optional regeneration bottle is fitted, connect the regeneration solution tubing QD coupling plug
E
to the mating
connector
F
at the base of base of the regeneration solution bottle.
4. Connect the calibration solution 1 tubing QD coupling plug
G
to the mating connector
H
at the base of the calibration
solution 1 bottle.
Fig. 3.5 Reagent, calibration and (optional) regeneration tubing connections
Navig
ator 50
0
Entrainment
T-piece
Calib Regeneration
valve assembly
To constant-head unit
To flowcell