Navigator 500
Sodium analyzer
6 Operation overview – transmitter
OI/ASO550–EN Rev. G
35
6.7.1 Calibration log entries
Possible calibration log entries along with a description are shown in Table 6.2. The possible
Audit Log
entries along with a
description are shown in Table 6.3. The
Diagnostic Log
shows the history of diagnostic messages that have been displayed in the
Diagnostic View
– see Appendix A, page 72 for diagnostic messages.
Log entry
Description
Cal Failed
Calibration procedure failed due to high / low slope or out of calibration solution or sample
temperature error during a calibration procedure.
Cal Aborted
Calibration aborted manually by the user.
Cal Missed
Scheduled calibration missed due to:
– user being in
Man. Valve Control
menu.
– manual calibration in progress (on same wet-section)
– user in
Service
login.
This log entry is also displayed if a calibration associated with a regeneration is missed due
to
No Sample
state.
Coefficients
(Slope and mV Offset)
The Slope and mV Offset are displayed – refer to Section 5.4, page 24 for a description of
these coefficients.
Table 6.2 Calibration Log entries
Log entry
Description
Power Failure
Power to the transmitter is lost.
Power Recovery
Transmitter restarted after a power loss.
In Config.
User in
Advanced / Configuration
mode.
Time / Date Changed
User has changed date / time.
Daylight Saving
Time changed due to daylight saving.
Table 6.3 Audit Log entries