![ABB Navigator 550 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/abb/navigator-550/navigator-550_operating-instructions-manual_278038.webp)
Navigator 500
Sodium analyzer
6 Operation overview – transmitter
36
OI/ASO550–EN Rev. G
6.8 Logging
Data recorded in the transmitter's internal memory can be
archived to a removable Secure Digital (SD) card or USB stick.
The transmitter continuously records
all
data to its internal
memory and keeps track of which data has been archived.
The amount of time that data remains in the transmitter's
internal memory depends on the sample rate – see Table 6.4.
Sample data is saved to removable media as comma-separated
files.
Configuration files are saved as binary-encoded files. Additional
files can also be archived:
Event log files (these files contain
Audit Log, Alarm Log,
Diagnostic Log
and
Calibration Log
data)
Data log files
Configuration files
The transmitter’s internal memory supports a maximum of 10
Data Log
and
Event Log
files only and a maximum of 8
Configuration
files. Durations for continuous recording are
shown in Table 6.4 (internal storage).
A 2GB SD card / USB stick has sufficient external storage
capacity for >5 years data.
6.8.1 SD card / USB stick
There are two methods of archiving to an SD card / USB stick:
An SD card / USB stick is kept in the transmitter
Data is archived to the SD card / USB stick automatically
at set intervals. Archiving continues until the SD card /
USB stick full; archiving then stops. To ensure all required
data is archived successfully, the SD card / USB stick
should be swapped periodically for an empty one.
It is advisable to back-up critical data stored on an SD
card / USB stick regularly. The transmitter's internal
memory provides a buffer for the most recent data only
so, if data stored on an SD / USB stick card is lost, it can
be re-archived.
Data is copied to an SD card / USB stick when
required
When a SD card / USB stick is inserted into the
transmitter, the media status can be set to
Online
causing
unarchived data to be copied to the media – see Section
7, page 42 /
Media Card
menu level.
Note.
ABB's DataManager software can be used to store
and view data archived from the transmitter.
5 s
10 s
30 s
1 m
5 / 10 / 30 m
1 hr
30
days
60
days
180
days
300
days
300
days
300
days
Table 6.4 Internal (flash) memory storage capacity
Caution.
To avoid potential damage or corruption to data
recorded on an SD card / USB stick, take care when
handling and storing. Do not expose to static electricity,
electrical noise or magnetic fields. When handling the SD
card / USB stick take care not to touch any exposed metal
contacts.
Note.
It is not advisable to leave a USB stick in the
transmitter permanently.