![ABB Navigator 550 Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/abb/navigator-550/navigator-550_operating-instructions-manual_278052.webp)
Navigator 500
Sodium analyzer
7 Menu descriptions
50
OI/ASO550–EN Rev. G
7.1.4 Display
Used to select the display language, setup
Operator
page templates (1 to 5), enable diagnostic,
view and log functions, set the device’s display brightness / contrast and set the time and date.
Menu
Display
Exit
Select
Menu
Comment
Default
Language
Selects the display language:
English / German / French / Italian / Spanish
.
English
Operator Templates
Enables a template to be assigned to a wet-section for display
purposes – refer to Section 6.5, page 31 for
Operator Template
examples.
Page 1
Page 1 (to 5) Template
Page 1
template cannot be set, this defaults to the number of
wet-sections connected.
Chart View
The chart can be configured to display the trend for 1, 2, 3 or 4 PV
and / or temperature values. The engineering ranges for the values
are configured in the
Sensor Setup
menu – see page 46.
Note
: This menu is displayed only when
Chart View
is
Enabled
at the
Operator Functions / Chart View
.
Channel 1 (to 4)
Source
Selects the signal (PV and temperature) to be displayed on the chart.
Tag
A 3-character, alphanumeric tag used to identify the parameter on
the chart.
Chart Duration
Selects the chart duration from 1, 2, 4, 8, 12, 16, 20 and 24 hours.
View/Log Enables
Enables / disables the following
View and Log
functions:
View
functions:
Diagnostics View / Signals View / Chart View / Alarm View / Analog
OP View
– refer to Section 6.6, page 33 for examples of
Operator
Pages
in
View
mode.
Log
functions:
Calibration Log / Alarm Log / Audit Log / Diagnostics Log –
refer to
Section 6.7, page 34 for examples of
Operator Pages
in
Log
mode.
Continued on next page…